EZ Series Phosphate Analysers

Overview

The EZ Series Online Analysers offer multiple options to monitor Phosphate in water. Typical applications are wastewater, drinking water, steam & power generation, surface water.

The EZ1000 Series Analysers use colorimetric analysis to measure Phosphate:
– EZ1031 Vanadate yellow method; standard measuring range: 0.1 – 10 mg/L PO₄-P
– EZ1032 Molybdate blue method; standard measuring range: 0.005 – 1 mg/L PO₄-P

Options for EZ1000 analysers include:
– Calibration to 10, 25 or 50% of standard range
– Internal dilution

Options for all analysers include:
– Multiple stream analysis (1-8 streams) reducing cost per sampling point
– Analogue and/or digital outputs for communication

Phosphate can also be measured in combination with Total Phosphorus. For this option please check the EZ7800 Series.

Some applications may require the use of a sampling/filtration or an external dilution system for sample preconditioning. For this option please check the EZ9000 Sample Preconditioning page.

Read more

Product information

Specifications: Total Phosphorus (TP) and/or Phosphate (PO4-P)

Measurement Method: Colorimetric measurement at a wavelength of 700 nm using ascorbic acid reduction solution and molybdate color reagent after persulfate digestion in an acidic medium, compliant with APHA 4500-P

Measurement Range

  • EZ782x: from 0.005 - 1 mg/L TP to 0.5 - 50 mg/L TP
  • EZ786x: from 0.005 - 1 mg/L TP/PO4-P to 0.5 - 50 mg/L TP/PO4-P

Accuracy: Better than 2% full scale for standard test solutions

Detection Limit: ≤ 0.005 mg/L

Measurement Cycle: Standard TP or TP and Orthophosphate: 30 minutes

Automatic Cleaning: Yes

Calibration: Automatic, 2 points; freely programmable frequency

Ambient Temperature: 10 - 30 °C ± 4 °C at relative humidity 5 - 95% (non-condensing)

Reagent Requirements: Keep between 10 - 30 °C (50 - 86 °F)

Sample Flow Rate: 100 - 300 mL/min

Sample Pressure: Atmospheric pressure overflow

Sample Quality: Maximum particle size 100 µm, <0.1 g/L; Turbidity <50 NTU Sample Temperature: 10 - 30 °C (50 - 86 °F)

Power Supply: 230 VAC, 50Hz

Service Gas: Dry and oil-free per ISA-S7.0.01-1996 quality standard

Deionized Water: For rinsing/dilution

Drain: Atmospheric pressure, vented, minimum Ø 64 mm

Grounding: Yes

Analog Output: 4-20 mA, Standard 1, maximum 8

Digital Output: Optional Modbus RS485/ Modbus TCP/IP

Alarms: 1 x fault, 4 x user-configurable

Dimensions (H x W x D): 690 mm x 465 mm x 330 mm

Protection Class: Analyzer cabinet: IP55 / Panel PC: IP65

Material: Hinge part: heat-resistant ABS, door: plexiglass; Wall part: Galvanized steel, powder-coated

Number of Sample Streams: 1 stream (Option up to 8 sample streams)

Calibration Verification: Automatic; freely programmable frequency

Weight: 25 kg

Certification: CE compliant / UL certified