Feature Value
Measuring Range Analyzer without cation pump: 0.01 ppb - 10,000 ppb
Analyzer with cation pump: 0.01 ppb - 200 ppm
Repeatability <0.02 ppb or 1.5% of the reading value (whichever is greater) within ± 10°C (± 50°F)
Limit of Detection (LDO) 0.01 ppb
Response Time From 0.1 ppb to 10 ppb: T90 ≤ 3 minutes, T95 ≤ 4 minutes
From <1 ppb to 100 ppb: T90 <2 minutes, T95 <3 minutes (approximately 150 seconds)
Calibration Method Automatic, manual 1 or 2 point
Sample Preparation For non-cation pump applications: Cation: Di-isopropylamine (DIPA) (1 L / 90 days) at 25°C for a target sample pH of 10.5
For cation pump applications: DIPA (1 L / month) at 25°C for a target sample pH of 10.5
Number of Channels 1, 2, 4 with programmable sequence
Maximum Allowable Suspended Solids Concentration in Sample <2 NTU, oil-free, grease-free. For boiler sample type, install a 100 µm filter
Acidity <50 ppm, non-cation application <250 ppm, cation application
Sample Temperature 5 - 45°C (41 - 113°F)
Ambient Temperature 5-50°C (41 - 122°F)
Sample Pressure 0.2 - 6 bar (3 - 87 psi)
Sample Flow Rate 100 - 150 mL/min (6 - 9 L/h)
Inputs Sample inlet and sample outlet: 6 mm OD push-fit fitting for connecting plastic tubing. Chemical drain: 7/16 inch ID slip fitting for soft plastic tubing
Power Requirements 100 - 240 VAC; 50/60 Hz
Protection Rating IP65
Display 5.7" Color LCD
Analog Output 6 x 0/4-20mA
Relays 6
Digital Input 6
Materials Polyol enclosure, PC door, PC hinges and latches, 304/316 SST hardware
Dimensions Analyzer with enclosure: 681 mm x 452 mm x 335 mm
Analyzer without enclosure: 681 mm x 452 mm x 254 mm
Weight Analyzer with enclosure: 20 kg
Analyzer without enclosure: 14 kg
Maintenance Cycle 90 days; replenish electrolyte solution, reactivate and calibrate
Measurement Principle The NA5600sc Online Sodium Analyzer utilizes a pH-adjusted ion selective electrode measurement. Sample pH conditioning is necessary to minimize the influence of temperature or other ions on the sodium measurement. Temperature stabilization and compensation are ensured by using additional conditioning reagents as sample temperature and pH change. In the case of multi-channel analysis, an "intelligent" rinse sequence between channels ensures a minimum cycle time of 10 minutes and prevents cross-contamination of samples.